Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: TIMS
  • Manufacturer: MSTech Group
  • Country: China
  • Available: Active
All specifications

About

Short description

High-precision double-focusing thermal ionization mass spectrometer (TIMS) with fixed multicollectors and variable dispersion optics.

Specifications

High-precision double-focusing thermal ionization mass spectrometer (TIMS) with fixed multicollectors and variable dispersion optics.

  • Status: Active
  • Country: China
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable Isotopes/Elements: Lithium to uranium isotopes, including 6Li/7Li, 234U/238U, and other actinides
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Thermal ionization
  • Signal Stability / Relative Standard Deviation
    : ±30 ppm over 30 minutes
  • Inlet Types: filament sample turret with single and double filament options, optional preheat system
  • High Resolution
    : Yes
  • Multicollector
    : Yes
  • Collectors Qty
    : Up to 15 fixed-position Faraday collectors and up to 4 full-size discrete dynode electron multipliers (SEM)
  • Sensitivity
    : High sensitivity; 238U isotopic sensitivity: ≤2 ppm (without RPQ), ≤0.1 ppm (with RPQ lens)
  • Abundance Sensitivity
    : ≤0.1 ppm with RPQ lens for uranium isotopes
  • Dynamic Range
    : dynamic range from 0 to 50V
  • Detection Limits
    : Sub-ppm detection limits
  • Precision
    : Internal precision: <0.002% for 238U (U500 standard)
  • Detector Type
    : Faraday cups and Secondary Electron Multipliers (SEM) with RPQ lens for improved sensitivity
  • Possible Amplifiers
    : 15 amplifiers, noise ≤ 40 μVrms (10¹¹ Ω, 4-second integration time)
  • Sample Turret
    : 21-position automated cathode drum
  • Geometry Type
    : Nier-Johnson straight geometry with double focusing
  • Focusing Type
    : Variable dispersion zoom optics using dual quadrupole lenses
  • Magnetic Sector Radius/ Angle
    : Magnetic analyzer radius: 350 mm; 90° deflection
  • Electric Sector Radius/ Angle
    : Electrostatic analyzer radius: 250 mm
  • Vacuum System
    : Static vacuum system, UHV ≤ 5 × 10^-8 mbar in the ion source, ≤ 5 × 10^-9 mbar in the analyzer Oil-free vacuum system with forevacuum, turbomolecular pump, and nitrogen trap
  • Gas Supply: Optional nitrogen bleed system

Other company product