Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: Unknown
  • Manufacturer: Markes International
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

Continuous dual-trap thermal desorption system for VOC and SVOC monitoring in air. Provides cryogen-free operation with independent sampling channels and automated calibration. Designed for remote or unattended operation with high sensitivity and low maintenance.

Specifications

Continuous dual-trap thermal desorption system for VOC and SVOC monitoring in air. Provides cryogen-free operation with independent sampling channels and automated calibration. Designed for remote or unattended operation with high sensitivity and low maintenance.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Device Class (Function)
    : Special system
  • Technique Compatibility
    : GC, GC-MS
  • Brand Independence
    : Yes
  • Performance Metrics
    : Trap heating up to 50 °C/s (programmed 1-40 °C/s); tube oven 30-440 °C; flow 2-250 mL/min; cryogenic range -30 to 50 °C; fully inert paths; repeat analysis via re-collection;.
  • Power Specifications: 100-240 V AC, 50/60 Hz, 1200 W
  • Environmental Specifications: 15-30 °C; 5-95 % RH (non-condensing); altitude ≤ 2000 m
  • Physical Specifications: 45 × 32 × 52 cm; 25 kg