Specifications

  • Category: Spectroscopy
  • Technique: LSC
  • Manufacturer: Revvity
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

High-performance time-resolved liquid scintillation counter for ultra-low-level and mixed radionuclide detection. Uses advanced PSA and dual-phase separation for superior sensitivity. Ideal for research, pharmaceutical, and environmental applications.

Specifications

High-performance time-resolved liquid scintillation counter for ultra-low-level and mixed radionuclide detection. Uses advanced PSA and dual-phase separation for superior sensitivity. Ideal for research, pharmaceutical, and environmental applications.

  • Status: Active
  • Country: United States of America
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Radioactive isotopes
  • Measurable Isotopes/Elements: ³H, ¹⁴C, ⁹⁰Sr, gross alpha/beta, and others
  • Sensitivity
    : Enhanced by dynamic color correction and luminescence adjustment
  • Background
    : Low background noise using external and internal corrections
  • Sample Capacity: Up to 408 × 20 mL or 720 × 4/7 mL vials with Varisette™ cassette system
  • Detection Type
    : Photomultiplier-based coincidence detection with dual PSA discriminator
  • Energy Range
    : 0–2000 keV
  • Efficiency Normal Count Mode
    : ³H: >63%, ¹⁴C: >95% (minimum acceptable efficiency)
  • Figure Of Merit (FOM (E2/B)
    : ³H: >180, ¹⁴C: >360 (normal mode); ³H: >500, ¹⁴C: >940 (ultra-low-level mode)
  • Observed Performance
    : Low background noise: ³H: 17 CPM, ¹⁴C: 26 CPM (average)
  • Alpha/Beta Separation
    : Yes, with PSA histogram for mixed sample resolution
  • Gas Supply: Not required
  • Power Specifications: 100–240 VAC, 50/60 Hz; Power consumption: <800 VA
  • Environmental Specifications: Operating temperature: 15–32°C; Humidity: 30–80%
  • Physical Specifications: 103 cm W × 47 cm H × 81 cm D; Weight: 216 kg (477 lbs)

Other company product