Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Nu Instruments
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

Collision Cell Multi-Collector Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Incorporates a high-energy ion path and a low-energy path with a collision/reaction cell.

Specifications

Collision Cell Multi-Collector Inductively Coupled Plasma Mass Spectrometer (ICP-MS) Incorporates a high-energy ion path and a low-energy path with a collision/reaction cell.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Mg, Si, S, K, Ca, Ti, V, Cr, Fe, Ni, Se
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Inductively Coupled Plasma
  • Inlet Types: Supports wet plasma, desolvation, hydride/cold vapor generation, and laser ablation
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : High resolution and pseudo high-resolution capabilities
  • Multicollector
    : Yes
  • Collectors Qty
    : 16 Faraday detectors, optional up to 6 ion-counting detectors (SEM/Daly)
  • Sensitivity
    : Enhanced sensitivity interface as standard
  • Abundance Sensitivity
    : High abundance sensitivity with deceleration filters
  • Dynamic Range
    : 0-75 Gcps (0-1200 V) from SEM to 10¹⁰ Ω resistor pre-amplifier
  • Detector Type
    : Faraday detectors, SEM/Daly detectors
  • Possible Amplifiers
    : Software-switchable resistor amplifiers
  • Reaction Cell System Details: Hexapole collision/reaction cell; up to 4 independently controlled reaction gas inlets; removal of isobaric interferences using charge exchange reactions
  • Spray Chamber
    : Peltier-cooled spray chamber
  • Nebulizer
    : Concentric nebulizer, compatible with desolvating nebulizers
  • Peristaltic Pump
    : Multi-channel peristaltic pump
  • Torch
    : High-performance torch for stability
  • Interface
    : Enhanced Sensitivity Interface
  • Geometry Type
    : Double focusing
  • Focusing Type
    : Variable Zoom Optics
  • Cooling System
    : Water-cooled system for optimal thermal stability
  • Vacuum System
    : Fully dry pumped configuration
  • Gas Supply: High-purity argon, helium, and optional custom gases for specific systems; Up to 4 mass flow-controlled collision/reaction gas inlets
  • Reaction Cell System: Yes