Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired High-Resolution Orbitrap GC-MS for Comprehensive Analysis.

Specifications

Retired High-Resolution Orbitrap GC-MS for Comprehensive Analysis.

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Orbitrap MS
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Up to 100,000 FWHM
  • Sensitivity
    : IDL ≤6 fg OFN
  • Dynamic Range
    : >10⁶
  • Available Configurations: EI, PCI, NCI
  • Scan Modes
    : Full Scan, SIM, tSIM, MS/MS
  • Ionization Modes [GC-MS]
    : Electron (EI), Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass Accuracy/Stability
    : <1 ppm
  • Scanning Speed
    : Up to 18 Hz
  • Direct Inlet Option
    : Yes
  • Ion Source Temperature
    : 50 to 350°C
  • Collision Cell
    : HCD Cell
  • Installation Specifications
    : EI IDL ≤6 fg OFN S/N ≥10,000:1 at 50,000 FWHM
  • Vacuum System
    : 2 Turbo pumps
  • Power Specifications: 2x230V, 15A, 50/60Hz
  • Environmental Specifications: 15-26°C, 40-70% RH
  • Physical Specifications: 95x91x148 cm, 254 kg