Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Nu Instruments
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

A third-generation multi-collector ICP-MS with Enhanced Abundance Transfer Optics (EATO) for improved sensitivity and a robust RF generator with frequency tuning.

Specifications

A third-generation multi-collector ICP-MS with Enhanced Abundance Transfer Optics (EATO) for improved sensitivity and a robust RF generator with frequency tuning.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Lithium to uranium, including actinides and heavy metals
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : <0.05 ppm with EATO and deceleration filters
  • Inlet Types: Laser ablation, desolvation systems, autosamplers, gas chromatography, and cold vapor generation
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : >800 (typical), with pseudo-high-resolution and full high-resolution capabilities
  • Multicollector
    : Yes
  • Collectors Qty
    : 16 Faraday cups (basic) with up to 6 ion counters (maximum)
  • Sensitivity
    : Dry plasma: U: 410 V/ppm, Pb: 500 V/ppm Wet plasma: U: 40 V/ppm, Pb: 50 V/ppm
  • Abundance Sensitivity
    : <2ppm at -1 amu spacing relative to 238U; <0.5 ppm with deceleration filter
  • Dynamic Range
    : >10¹² with enhanced collector and amplifier configurations
  • Detection Limits
    : Sub-ppq for trace isotopes
  • Precision
    : High precision for isotopic ratio analysis; <0.032% RSD for Nd isotopes over 7 hours
  • Detector Type
    : Faraday detectors, discrete dynode SEMs, and Daly detectors
  • Possible Amplifiers
    : 10¹⁰, 10¹¹, 10¹², and 10¹³ ohm resistors with software-switchable configurations
  • Spray Chamber
    : Peltier-cooled cyclonic spray chamber
  • Nebulizer
    : Concentric nebulizer
  • Peristaltic Pump
    : multi-channel for sample introduction and drainage
  • Torch
    : Advanced torch box design for flexibility and stability
  • Interface
    : High-capacity Jet Interface with optimized pumping for vacuum stability
  • Geometry Type
    : Nier-Johnson forward geometry
  • Focusing Type
    : Double focusing with magnetic and electrostatic sectors
  • Magnetic Sector Radius/ Angle
    : 250 mm/70 deg
  • Electric Sector Radius/ Angle
    : 350 mm/ 70 deg
  • Cooling System
    : Water-cooled system for optimal thermal stability
  • Vacuum System
    : Fully turbo-pumped with additional rotary pumps for plasma and interface regions
  • Gas Supply: Argon with additional helium for laser ablation systems
  • Reaction Cell System: No