Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: LECO
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

High-Resolution TOF-GC-MS system with Dual flight paths.

Specifications

High-Resolution TOF-GC-MS system with Dual flight paths.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Time-of-Flight MS
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : HR: 25,000; amuHR: 50,000
  • Sensitivity
    : IDL 100 fg OFN for HR-EI installation
  • Dynamic Range
    : 4 orders
  • Available Configurations: HR EI, HR CI optional
  • Ionization Modes [GC-MS]
    : Electron (EI), Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass Accuracy/Stability
    : <1 ppm
  • Detector Type [GC-MS]
    : Folded Flight Path TOF (FFP) with Long Life MCP Detector
  • Scanning Speed
    : Up to 200 spectra/sec
  • Installation Specifications
    : IDL <100 fg OFN HR-EI mode
  • Vacuum System
    : 2 Turbo pumps
  • Physical Specifications: 130x125x83 cm, 431 kg