Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: LECO
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

High throughput Benchtop TOF-GC-MS with StayClean Ion Source.

Specifications

High throughput Benchtop TOF-GC-MS with StayClean Ion Source.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Time-of-Flight MS
  • Mass Resolution/ Mass Resolving Power
    : ≥1500 for m/z 1500, ≥1100 for m/z 219, Mass peak width ≤ 0.2 Da FWHH for m/z 219
  • Sensitivity
    : S/N ≥10 for 50 fg OFN on-column, IDL ≤20 fg OFN
  • Dynamic Range
    : 10⁵ (Linear)
  • Available Configurations: EI
  • Scan Modes
    : OFN NIST score >800 on 1 pg OFN
  • Ionization Modes [GC-MS]
    : Electron (EI)
  • Mass Accuracy/Stability
    : <0.30 Da for entire mass range, <0.30 Da drift for 24 hours without calibration
  • Detector Type [GC-MS]
    : Time-of-Flight (TOF) with single mirror, dual stage reflectron
  • Scanning Speed
    : 1 to 50 spectra/sec (35,000 native spectra/sec maximum)
  • Ion Source Temperature
    : 100°C to 300°C
  • Maximum Column Flow Rate
    : Helium, 1 mL/min recommended, 5 mL/min maximum
  • Installation Specifications
    : ≤20 fg OFN, Precision ≤5% RSD for 1 pg OFN
  • Environmental Specifications: Operating Temperature 16°C to 26°C, Storage 0°C to 50°C
  • Physical Specifications: 50.8 cm (H) × 53.3 cm (W) × 81.3 cm (D), ~91 kg

Other company product