Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Tribrid MS with Orbitrap, quadrupole, linear ion trap; high sensitivity; multiple fragmentation modes

Specifications

Tribrid MS with Orbitrap, quadrupole, linear ion trap; high sensitivity; multiple fragmentation modes

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Quadrupole MS (single), Orbitrap MS, Ion Trap MS
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : 500,000 FWHM at m/z 200
  • Sensitivity
    : 1 pg reserpine S/N 100:1
  • Dynamic Range
    : 4 orders of magnitude
  • Scan Modes
    : Full MS, MS/MS, MSn, SIM, tMSn, targeted MS2
  • Ion Source Types [LC-MS]
    : EASY-Spray, Nanospray Flex, HESI-II, APCI
  • Mass Accuracy/Stability
    : <1-3 ppm depending on calibration type
  • Scanning Speed
    : Orbitrap MSn up to 20 Hz; ion trap MSn up to 20 Hz
  • Polarity Switching
    : 1 sec or less
  • Fragmentation
    : CID, HCD, ETD
  • Vacuum System
    : Split-flow turbo pump
  • Power Specifications: 200-240 Vac, 15 A; ~3,500 W
  • Environmental Specifications: 15-26 °C; 50-80% RH
  • Physical Specifications: 127.0 × 76.7 × 70.3 cm, 180 kg

Other company product