Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Tribrid quadrupole-linear ion trap-Orbitrap; High-field Orbitrap analyzer; Multiple fragmentation capabilities

Specifications

Tribrid quadrupole-linear ion trap-Orbitrap; High-field Orbitrap analyzer; Multiple fragmentation capabilities

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Quadrupole MS (single)
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Up to 480,000 FWHM at m/z 200
  • Dynamic Range
    : 5 orders of magnitude (eDR mode)
  • Scan Modes
    : Full MS, SIM, MS/MS, PRM, DIA, AIF, MSⁿ, dd-MSⁿ, tSIM
  • Ion Source Types [LC-MS]
    : Nanospray Flex, EASY-Spray NG, Ion Max NG with HESI, APCI; FAIMS Pro Duo
  • Mass Accuracy/Stability
    : <1 ppm RMS (internal), <3 ppm RMS (external)
  • Scanning Speed
    : Up to 45 Hz (Orbitrap); Ion trap up to 66 Hz
  • Polarity Switching
    : 30 ms
  • Fragmentation
    : CID, HCD, ETD, EThcD
  • Vacuum System
    : Multi-stage with turbomolecular + dry pumps
  • Power Specifications: 200-240 V, 50/60 Hz, 30 A
  • Environmental Specifications: 18-27 °C; 20-80% RH
  • Physical Specifications: 142.0 × 87.0 × 142.0 cm; 650.0 kg

Other company product