Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Noble Gas MS
  • Manufacturer: Isotopx
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

Advanced noble gas mass spectrometer with high precision for isotope ratio analysis of noble gases. Features include ATONA® amplifiers, low-volume static vacuum analyzer, and large-radius magnet.

Specifications

Advanced noble gas mass spectrometer with high precision for isotope ratio analysis of noble gases. Features include ATONA® amplifiers, low-volume static vacuum analyzer, and large-radius magnet.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Noble gases
  • Measurable Isotopes/Elements: Isotopes of noble gases
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Electron impact
  • Signal Stability / Relative Standard Deviation
    : <25 ppm drift over 30 minutes
  • Inlet Types: Fully automated inlet system with cryogenic traps, getters, and high-vacuum components
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : >600 (10% valley definition)
  • Multicollector
    : Yes
  • Collectors Qty
    : 10: 9 Faraday cups and 1 secondary electron multipliers ((optional more SEM)
  • Sensitivity
    : Helium: 2×10−4 A/Torr; Argon: 1×10−3 A/Torr
  • Abundance Sensitivity
    : <1 ppm< for 40Arr at 39A
  • Dynamic Range
    : 1nA to 1aA
  • Detection Limits
    : Sub-ppb levels for noble gases
  • Precision
    : <0.01%for isotopic ratios
  • Detector Type
    : Faraday detectors with ATONA® amplifiers and discrete dynode SEMs
  • Possible Amplifiers
    : ATONA® amplifier (equivalent headroom of 100V measured on a 10¹¹Ω resistor with the equivalent noise characteristic of a 1014Ω amplifier at 20 secs integration)
  • Background
    : 5×10−14 cc STP at mass 36Ar
  • Geometry Type
    : Double-focusing with extended geometry
  • Focusing Type
    : Single focusing
  • Magnetic Sector Radius/ Angle
    : 27 cm radius; 90° deflection
  • Cooling System
    : Liquid nitrogen cryotrap and air-cooled electronics
  • Vacuum System
    : Turbo molecular pumps (80 L/s), ion-getter pump (40 L/s), static vacuum chamber
  • Gas Supply: High-purity noble gases

Other company product