Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Inactive
All specifications

About

Short description

Retired multicollector ICP-MS optimized for isotopic and elemental analysis.

Specifications

Retired multicollector ICP-MS optimized for isotopic and elemental analysis.

  • Status: Inactive
  • Country: Germany
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: U, Pb, Hf, Nd, Sr, and more, including Li isotopes and actinides
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : <2% over one hour
  • Inlet Types: Laser ablation, gas chromatography, liquid chromatography, autosamplers
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Up to 10,000
  • Multicollector
    : Yes
  • Collectors Qty
    : Configurable up to 9 Faraday cups (and 8 ion counters)
  • Sensitivity
    : Standard: 7Li: 10 V/ppm; 208Pb: 20 V/ppm; 238U: 40 V/ppm With desolvation: 7Li: 60 V/ppm; 208Pb: 100 V/ppm; 238U: 200 V/ppm
  • Abundance Sensitivity
    : <5 ppm for 237/238 without RPQ; <500 ppb with RPQ (transmission >90%)
  • Dynamic Range
    : Up to 10¹²
  • Detection Limits
    : Sub-ppq levels for specific isotopes (e.g., uranium)
  • Precision
    : Sub-ppm precision for isotopic ratios
  • Detector Type
    : Faraday cups and discrete dynode electron multipliers
  • Possible Amplifiers
    : Virtual amplifiers with selectable 10¹⁰, 10¹¹, and 10¹² ohm resistors
  • Reaction Cell System Details: RPQ lens for isobaric interference reduction with high transmission
  • High Matrix Introduction System (HMI)
    : optional
  • Spray Chamber
    : Peltier-cooled cyclonic spray chamber
  • Nebulizer
    : Concentric nebulizer
  • Peristaltic Pump
    : Not explicitly mentioned
  • Torch
    : Torch with platinum-tipped guard electrode
  • Interface
    : Jet Interface with enhanced pumping capacity
  • Geometry Type
    : Nier-Johnson geometry
  • Focusing Type
    : Double focusing with magnetic and electrostatic sectors
  • Magnetic Sector Radius/ Angle
    : 230 mm/ 90 deg
  • Electric Sector Radius/ Angle
    : 220 mm/ 90 deg
  • Cooling System
    : Water-cooled
  • Vacuum System
    : Fully turbo-pumped system with enhanced capacity for Jet Interface
  • Gas Supply: High-purity argon and optional collision/reaction gases for RPQ lens
  • Reaction Cell System: Yes
  • Environmental Specifications: Temperature: 18–24°C; humidity: 50–60%, non-condensing
  • Physical Specifications: 2450х 1890х 1740, 2200 kg

Other company product