Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Multicollector ICP-MS with MS/MS mode for removing isobaric interferences and enhancing accuracy in isotopic measurements.

Specifications

Multicollector ICP-MS with MS/MS mode for removing isobaric interferences and enhancing accuracy in isotopic measurements.

  • Status: Active
  • Country: Germany
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Rare earth elements, actinides, and transition metals
  • Analyzer Type
    : Magnetic Sector MS, Tandem Magnetic Sector MS
  • Ion Source Type: Inductively Coupled Plasma
  • Inlet Types: Laser ablation, autosamplers, HPLC, GC
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Low, medium, high, and extra high resolution (>15,000)
  • Multicollector
    : Yes
  • Collectors Qty
    : 10 moveable Faraday cups and 1 central dual-mode detector (Faraday/SEM)
  • Abundance Sensitivity
    : <50 ppb abundance sensitivity at 1 amu
  • Dynamic Range
    : Over 9 orders of magnitude (up to 60 Gcps)
  • Detection Limits
    : Sub-ppq levels with enhanced interference removal via MS/MS
  • Precision
    : High precision for isotopic and elemental ratio analysis, especially with MS/MS
  • Detector Type
    : Faraday cups and discrete dynode electron multiplier
  • Possible Amplifiers
    : 24 amplifier slots, 12 x 10¹¹ Ω amplifiers, 4 x 10¹³ Ω amplifiers
  • Reaction Cell System Details: MS/MS with pre-cell quadrupole filtering
  • Spray Chamber
    : Peltier-cooled cyclonic spray chamber
  • Nebulizer
    : Concentric nebulizer
  • Torch
    : High-performance torch system
  • Interface
    : Jet interface optional
  • Geometry Type
    : Nier-Johnson geometry with MS/MS addition
  • Focusing Type
    : Double focusing with additional MS/MS quadrupole focusing
  • Magnetic Sector Radius/ Angle
    : Large-radius
  • Cooling System
    : Water-cooled
  • Vacuum System
    : Fully turbo-pumped
  • Gas Supply: High-purity argon (>99.996%) with reactive gases (O2, N2O, NH3, SF6) and collision gases (H2, He)
  • Reaction Cell System: Yes