Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: TIMS
  • Manufacturer: Uralpribor
  • Country: Russian Federation
  • Available: Inactive
All specifications

About

Short description

High-precision thermal ionization mass spectrometer (TIMS) designed for isotopic analysis of uranium in solid-phase samples. Provides automated sample preparation, high resolution, and robust performance for nuclear and geochemical applications.

Specifications

High-precision thermal ionization mass spectrometer (TIMS) designed for isotopic analysis of uranium in solid-phase samples. Provides automated sample preparation, high resolution, and robust performance for nuclear and geochemical applications.

  • Status: Inactive
  • Country: Russian Federation
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Thermoionizing elements (Sr, Nd, Os, Pb, U, and others)
  • Measurable Isotopes/Elements: 234U, 235U, 236U, 238U
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Thermal ionization
  • Signal Stability / Relative Standard Deviation
    : ≤ 20 ppm over 20 minutes for 238U
  • Inlet Types: Filament loading system with manual or automated adjustment
  • Mass Resolution/ Mass Resolving Power
    : 750
  • Multicollector
    : Yes
  • Collectors Qty
    : 10
  • Sensitivity
    : optimized for trace-level detection uranium isotopes
  • Abundance Sensitivity
    : ≤ 5 ppm for 238U
  • Precision
    : ±0.04% for 235U in the range 0.5%−1.0% for 234U/236U at 0.0005%−0.006%
  • Detector Type
    : Faraday cups and secondary electron multiplier (SEM)
  • Possible Amplifiers
    : High-stabilitHighly sensitive input amplifier in a special vacuum chamber with suppressed microphone effect, ensuring measurement accuracy.y amplifiers with reduced noise for ultra-precise measurements
  • Geometry Type
    : Double-focusing geometry with magnetic sector
  • Focusing Type
    : Electrostatic focusing system
  • Power Specifications: 220 V ±10%, 50 Hz; Maximum power consumption: 3 kVA
  • Environmental Specifications: Temperature: 10–35°C; Humidity: ≤ 80%; Atmospheric pressure: 84–106.7 kPa
  • Physical Specifications: Analytical unit: 1800×1450×1800 mm; Electronics unit: 1200×650×1700 mm; Operator desk: 1500×1250×850 mm; Weight: 1400 kg

Other company product