Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: GD-MS
  • Manufacturer: Lumex
  • Country: Russian Federation
  • Available: Active
All specifications

About

Short description

Time-of-Flight Mass Spectrometry (TOF-MS) with glow discharge plasma ionization. Real-time analysis of volatile organic and inorganic compounds in gas phase.

Specifications

Time-of-Flight Mass Spectrometry (TOF-MS) with glow discharge plasma ionization. Real-time analysis of volatile organic and inorganic compounds in gas phase.

  • Status: Active
  • Country: Russian Federation
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Most of elements, isotopes, and molecular species
  • Measurable Isotopes/Elements: Isotopic and molecular composition of gas-phase compounds
  • Analyzer Type
    : Time-of-Flight MS
  • Ion Source Type: Glow discharge plasma (Penning ionization), optional electron ionization
  • Inlet Types: Direct gas introduction
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Up to 6000 (m/Δm at FWHM); nominal 4000
  • Sensitivity
    : Down to 10–100 ppt
  • Detection Limits
    : 10–100 ppt
  • Discharge Parameters
    : Glow discharge in air, N₂, Ar, He, Ne, etc.
  • Geometry Type
    : TOF with reflectron
  • Focusing Type
    : Gridless ion mirror (reflectron focusing)
  • Vacuum System
    : One forepump and two turbo molecular pumps (240 L/s)
  • Gas Supply: Air, N₂, Ar, He, Ne, others
  • Physical Specifications: 1450 × 780 × 1550 mm

Other company product