Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Single quadrupole IC/LC-MS. Dual-role design. Routine assays.

Specifications

Single quadrupole IC/LC-MS. Dual-role design. Routine assays.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Quadrupole MS (single)
  • High Resolution
    : No
  • Mass Resolution/ Mass Resolving Power
    : Unit mass resolution
  • Sensitivity
    : HESI pos: 10 pg Reserpine 400:1; HESI neg: 20 pg p-Nitrophenol 500:1
  • Dynamic Range
    : 7 orders of magnitude
  • Scan Modes
    : Full Scan, SIM, simultaneous Full Scan/SIM
  • Ion Source Types [LC-MS]
    : HESI-II
  • Mass Accuracy/Stability
    : ≤ ±0.1 Da; better than 0.1 Da over 48 h
  • Scanning Speed
    : Up to 20,000 Da/s
  • Polarity Switching
    : 25 ms
  • Vacuum System
    : External roughing pump; API source vacuum interlock
  • Power Specifications: 100-240 VAC, 50/60 Hz
  • Environmental Specifications: 15-35 °C; 20-80% RH
  • Physical Specifications: 52.0 × 42.0 × 91.0 cm; 71.0 kg

Other company product