Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: GV Instruments
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Inactive
All specifications

About

Short description

Retired multicollector ICP mass spectrometer equipped with a hexapole collision cell and advanced detector array.

Specifications

Retired multicollector ICP mass spectrometer equipped with a hexapole collision cell and advanced detector array.

  • Status: Inactive
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Includes rare earths, transition metals, and actinides
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Inductively Coupled Plasma
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Up to 10,000
  • Multicollector
    : Yes
  • Collectors Qty
    : 9 Faraday cups + 7 ion counting detectors (optional)
  • Abundance Sensitivity
    : <5ppm (<50 ppb with WARP filter)
  • Detection Limits
    : Sub-ppm detection limits
  • Precision
    : Isotope ratio precision for Ag-107/Ag-109: < 0.03%
  • Detector Type
    : Daly detector and conversion dynode multipliers
  • Possible Amplifiers
    : Variable gain amplifiers
  • Reaction Cell System Details: Hexapole collision cell
  • Cooling System
    : Water-cooled magnet
  • Vacuum System
    : Ion pumps
  • Reaction Cell System: Yes

Other company product

Loading other products...