Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Shimadzu
  • Country: Japan
  • Available: Active
All specifications

About

Short description

High-sensitivity ICP-MS with advanced mini-torch and reaction mode for enhanced interference removal.

Specifications

High-sensitivity ICP-MS with advanced mini-torch and reaction mode for enhanced interference removal.

  • Status: Active
  • Country: Japan
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Analyzer Type
    : Quadrupole MS (single)
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : <1% RSD short-term; <2% RSD long-term
  • Mass Resolution/ Mass Resolving Power
    : 0.5 u for high-resolution mode
  • Dynamic Range
    : 8 orders of magnitude
  • Detector Type
    : Discrete dynode electron multiplier
  • Reaction Cell System Details: Octopole collision/reaction cell with helium and other reactive gases (e.g., H2)
  • Spray Chamber
    : Thermoelectrically cooled cyclone chamber
  • Nebulizer
    : High-efficiency concentric nebulizer
  • Peristaltic Pump
    : mini-pump
  • Torch
    : Mini-torch design for lower argon consumption
  • Interface
    : Tool-free interface for quick maintenance
  • Focusing Type
    : RAPID (Right Angular Positive Ion Deflection) lens for enhanced ion transmission
  • Cooling System
    : External water circulator required
  • Vacuum System
    : Three-stage differential pumping with split-flow turbomolecular pump
  • Gas Supply: Low-purity argon gas (99.95%) compatible
  • Reaction Cell System: Yes
  • Power Specifications: 200–240 V ± 10 %, 50/60 Hz, 6 kVA
  • Environmental Specifications: Temperature: 18–28 °C; Humidity: 20–70% non-condensing
  • Physical Specifications: 853 mm (width) x 660 mm (depth) x 554 mm (height); Weight: 144 kg