Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: IRMS
  • Manufacturer: Sercon
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

Small-radius isotope ratio mass spectrometer with dual inlet and continuous flow modes.

Specifications

Small-radius isotope ratio mass spectrometer with dual inlet and continuous flow modes.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Light stable isotopes (C, N, S, O, H)
  • Measurable Isotopes/Elements: H₂, N₂, NO, N₂O, O₂, CO, CO₂, SO, SO₂
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Electron impact
  • Signal Stability / Relative Standard Deviation
    : <0.03 ppm/nA/hour
  • Inlet Types: Dual inlet, continuous flow, and microvolume systems
  • Mass Resolution/ Mass Resolving Power
    : 110 (m/Δm, 10% valley) for N₂, 40 for H₂
  • Multicollector
    : Yes
  • Collectors Qty
    : Universal triple collectors for simultaneous adjacent masses (28, 29, 30 to 64, 65, 66); dedicated HD detector for m/z 3
  • Sensitivity
    : Dual Inlet: <650 molecules CO₂/mass 44 ion; Continuous Flow: <850 molecules CO₂/mass 44 ion
  • Abundance Sensitivity
    : <1 ppm for H₂; <5 ppm for N₂; <30 ppm for CO₂ in continuous flow mode; <3 ppm for CO₂ in dual inlet mode
  • Detection Limits
    : Sub-ppm levels, depending on the isotopic system and application
  • Precision
    : High precision for isotope ratio measurements; internal precision up to 0.01‰ for CO₂ (¹³C and ¹⁸O)
  • Detector Type
    : High-efficiency Faraday collectors
  • Possible Amplifiers
    : Variable-gain amplifiers with software switchable options
  • H3 Factor
    : <5 ppm/nA
  • Geometry Type
    : 120° extended magnetic sector geometry
  • Focusing Type
    : Double stigmatic focusing for enhanced ion beam precision
  • Magnetic Sector Radius/ Angle
    : 11 cm radius; effective 21 cm radius dispersion; additional long spur with 98.8° sector
  • Cooling System
    : Integrated system for maintaining thermal stability
  • Vacuum System
    : Differentially pumped by drag-stage turbomolecular pumps with backing rotary pump; achieves ultra-high vacuum conditions
  • Gas Supply: Supports He, H₂, N₂, CO₂, and other high-purity gases; compatible with continuous flow applications
  • Power Specifications: 100-240 VAC; supports global power standards
  • Physical Specifications: 70 cm x 82 cm x 80 cm