Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Hewlett-Packard
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired benchtop ICP-MS with easy-to-use software, designed for environmental and routine metal analysis.

Specifications

Retired benchtop ICP-MS with easy-to-use software, designed for environmental and routine metal analysis.

  • Status: Inactive
  • Country: United States of America
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Analyzer Type
    : Quadrupole MS (single)
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : Short-term RSD < 1% (10 replicates); Long-term RSD < 2% (4 hours)
  • Mass Resolution/ Mass Resolving Power
    : 0.3–2.0 amu
  • Sensitivity
    : >8 Mcps/ppm for Li; >12 Mcps/ppm for Y, Tl
  • Dynamic Range
    : 8 orders of magnitude
  • Detection Limits
    : <1 ng/L for 51 elements; 1–10 ng/L for 18 elements, including Fe
  • Detector Type
    : Discrete dynode electron multiplier
  • Oxides %
    : CeO/Ce < 0.005
  • Spray Chamber
    : Peltier-cooled spray chamber
  • Nebulizer
    : V-groove nebulizer
  • Peristaltic Pump
    : 12-roller, low-pulsation pump
  • Torch
    : Demountable quartz torch with computer-controlled xyz translation stage
  • Cooling System
    : Water chiller or water recirculator required
  • Vacuum System
    : 3-stage system with one rotary pump and two turbomolecular pumps
  • Gas Supply: Argon
  • Reaction Cell System: No
  • Physical Specifications: 110 cm (width) x 60 cm (depth) x 58 cm (height); Weight: 135 kg

Other company product