Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: Noble Gas MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: Germany
  • Available: Active
All specifications

About

Short description

High-resolution multicollector static vacuum mass spectrometer (SVMS) designed for noble gas analysis. Equipped with a 35 cm, 120° magnetic sector analyzer and advanced ion optics for precision isotopic measurements of small gas samples.

Specifications

High-resolution multicollector static vacuum mass spectrometer (SVMS) designed for noble gas analysis. Equipped with a 35 cm, 120° magnetic sector analyzer and advanced ion optics for precision isotopic measurements of small gas samples.

  • Status: Active
  • Country: Germany
  • Isotopic: Yes
  • Analyte Type / Target Of Analysis: Noble gases
  • Measurable Isotopes/Elements: Isotopes of noble gases
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Electron impact
  • Signal Stability / Relative Standard Deviation
    : <0.2% drift per hour at 100 KCPS; Drift of ±25 ppm/30 min
  • Inlet Types: Dual inlet with getter pumps, cold traps, and pipette systems for precise sample handling
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Standard resolution: >600; High resolution: >1500 (10% peak valley definition)/ Res. Power > 5500 @ 0.05 mm source slit.Typical achieved > 7000
  • Multicollector
    : Yes
  • Collectors Qty
    : 5 Combined Faraday-Multiplier (CFM) detectors, including high-resolution options
  • Sensitivity
    : Helium: >2×10−4 Amps/Torr; Argon: >1×10−3 Amps/Torr
  • Abundance Sensitivity
    : < 1 ppm for adjacent masses sensitivity (39 contribution from 40)at a pressure of 1 x 10-7 mbar
  • Dynamic Range
    : 2 cps to ~1.4 × 10⁶ cps
  • Detection Limits
    : Sub-ppb levels for noble gas isotopes
  • Precision
    : <0.01% for isotopic ratios of noble gases
  • Detector Type
    : Faraday cups with suppressed noise and compact discrete dynode (CDD) electron multipliers
  • Possible Amplifiers
    : 10¹⁰, 10¹¹, 10¹², and 10¹³ Ω resistors
  • Background
    : ≤ 5e-13 cc STP at mass 36; Dark noise: <10 CPM (0.2 CPM typical)
  • Geometry Type
    : Extended geometry with 120° deflection
  • Focusing Type
    : Single focusing
  • Magnetic Sector Radius/ Angle
    : 35 cm radius, 120° deflection
  • Cooling System
    : Liquid nitrogen cooling for getters; water cooling for turbo pumps
  • Vacuum System
    : Ultra-high static vacuum system with turbo molecular and ion-getter pumps
  • Gas Supply: High-purity noble gases (e.g., helium, neon, argon)