Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: GD-MS
  • Manufacturer: MSI (Mass Spectrometry Instruments)
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Active
All specifications

About

Short description

High-resolution glow discharge mass spectrometer (GD-MS) for ultra-trace elemental and isotopic analysis. Combines a proven optical platform, modern stable electronics, and advanced depth-profiling capabilities for coatings, composites, and high-purity materials.

Specifications

High-resolution glow discharge mass spectrometer (GD-MS) for ultra-trace elemental and isotopic analysis. Combines a proven optical platform, modern stable electronics, and advanced depth-profiling capabilities for coatings, composites, and high-purity materials.

  • Status: Active
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Isotopes and trace components in metals, alloys, semiconductors, and insulators
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Glow discharge
  • Inlet Types: Pin and flat sample holders for various geometries, including unconventional shapes
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Continuously variable high resolution capabilities from <400 to >10,000 RP (10% valley definition)
  • Dynamic Range
    : Up to 12 orders of magnitude
  • Detection Limits
    : Sub-ppb levels for trace impurities
  • Detector Type
    : Dual collectors: Faraday cup and secondary electron multiplier (SEM)
  • Geometry Type
    : Double-focusing magnetic sector with electrostatic and magnetic analyzers
  • Focusing Type
    : Laminated magnet for stable ion focusing
  • Vacuum System
    : High-capacity turbo molecular pumps and cryogenic cooling
  • Gas Supply: Argon