Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: EXPEC Technology
  • Country: China
  • Available: Active
All specifications

About

Short description

ICP mass-spectrometer with dual-mode detector, advanced gas handling, and matrix tolerance.

Specifications

ICP mass-spectrometer with dual-mode detector, advanced gas handling, and matrix tolerance.

  • Status: Active
  • Country: China
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Analyzer Type
    : Tandem Quadrupole MS
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : 2 hour stability 2%
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : < 0.3 amu
  • Sensitivity
    : Li (7Li) 1,2 x 108; Co (59Co) 2,5 x 108; Cd (114Cd) 3 x 107; Pb (206Pb) 4 x 107 (imp/s)/(mg/dm3)
  • Dynamic Range
    : >10 orders of magnitude
  • Precision
    : RSD 2%
  • Oxides %
    : (138Ва16О+/138Ва+) 0,5%
  • Reaction Cell System Details: Five-channel collision reaction gas with waterfall-flow distribution
  • High Matrix Introduction System (HMI)
    : Yes
  • Torch
    : Vertical, anti-salt accumulation
  • Injector
    : High salt tolerance and organic injection
  • CRC Gas Flow
    : up to 10 MFCs configurable
  • Focusing Type
    : Off-axis with right-angle ion optics
  • Cooling System
    : Water chiller or water recirculator required
  • Vacuum System
    : Three turbomolecular pumps with five-stage vacuum
  • Gas Supply: Integrated system with 5-channel gas handling
  • Reaction Cell System: Yes
  • Power Specifications: 220 V ±20 V, 50 Hz
  • Environmental Specifications: Temperature: 18-24 °C, Humidity: 20-60%
  • Physical Specifications: H 700 mm W 1200mm D 680mm 300 kg