Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: Bruker / Scion
  • Country: Netherlands
  • Available: Active
All specifications

About

Short description

Ultra-compact triple quadrupole GC-MS/MS. Features fast and easy methods development with Compound Based Scanning (CBS).

Specifications

Ultra-compact triple quadrupole GC-MS/MS. Features fast and easy methods development with Compound Based Scanning (CBS).

  • Status: Active
  • Country: Netherlands
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Tandem Quadrupole MS
  • Mass Resolution/ Mass Resolving Power
    : User adjustable 0.7 to 4 Da
  • Sensitivity
    : MRM: 100 fg OFN for m/z 272>222, S/N ≥ 5000:1
  • Dynamic Range
    : >10⁷
  • Available Configurations: EI, EI (CI optional), EI/CI
  • Scan Modes
    : Full Scan, Precursor, Product, Neutral Loss/Gain Monitoring, Selected Ion Monitoring (SIM), Multiple Reaction Monitoring (MRM) and Result Dependent Scanning
  • Ionization Modes [GC-MS]
    : Electron (EI), Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass Accuracy/Stability
    : ±0.1 Da over 24 hours
  • Detector Type [GC-MS]
    : EDR Electron Multiplier
  • Scanning Speed
    : Up to 20,000 Da/sec
  • Ion Source Temperature
    : 100 to 350°C
  • Maximum Column Flow Rate
    : Up to 25 mL/min
  • MRM Transition Rate
    : 500 MRM’s/sec
  • Collision Cell
    : 180° curved path with pre- and post-filter regions
  • Installation Specifications
    : EI MRM IDL 25 fg OFN (S/N ≥ 50:1); EI Full Scan 1 pg OFN (S/N ≥ 1000:1)
  • Vacuum System
    : Dual-stage turbomolecular pump, air-cooled
  • Power Specifications: 100-240V, 50/60 Hz, 1200 VA
  • Environmental Specifications: 15–33°C, 20–80% RH
  • Physical Specifications: 45×28×57 cm, 40 kg