Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: Thermo Fisher Scientific
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired single Quadrupole GC-MS with Curved Pre-filter and Off-axis optics.

Specifications

Retired single Quadrupole GC-MS with Curved Pre-filter and Off-axis optics.

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Quadrupole MS (single)
  • Mass Resolution/ Mass Resolving Power
    : Unit mass
  • Sensitivity
    : 1 pg OFN S/N ≥75-100
  • Available Configurations: EI with 70 L/s Turbo, EI/CI with 250 L/s Turbo, EI/CI with 200/200 L/s Split-flow Turbo
  • Scan Modes
    : Full Scan, SIM
  • Ionization Modes [GC-MS]
    : Electron (EI), Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass Accuracy/Stability
    : ±0.1 amu/48h
  • Detector Type [GC-MS]
    : Conversion Dynode & Electron Multiplier
  • Scanning Speed
    : Up to 10,000 amu/sec
  • Ion Source Temperature
    : 125-300°C
  • Installation Specifications
    : S/N 20 fg OFN in EI SIM
  • Vacuum System
    : Turbo Pump 70/250/200 L/s (depending on config)
  • Power Specifications: 120/230V, 50/60 Hz
  • Environmental Specifications: 15-31°C, 40-80% RH
  • Physical Specifications: 44x33x68 cm, 45 kg