Specifications

  • Category: Chromatography
  • Technique: Unknown
  • Manufacturer: KNAUER
  • Country: Germany
  • Available: Active
All specifications

About

Short description

Single-wavelength UV/VIS detector for routine and fast LC methods. Compact design with deuterium lamp and wavelength range of 190-750 nm. Flow cell accessible and interchangeable for analytical or preparative LC.

Specifications

Single-wavelength UV/VIS detector for routine and fast LC methods. Compact design with deuterium lamp and wavelength range of 190-750 nm. Flow cell accessible and interchangeable for analytical or preparative LC.

  • Status: Active
  • Country: Germany
  • Device Class (Function)
    : Optical / Spectroscopic Detector
  • Technique Compatibility
    : LC
  • Brand Independence
    : Yes
  • Performance Metrics
    : Detector type - variable single wavelength UV detector; channel 1; light source D₂ lamp; range 190-750 nm; bandwidth 11 nm; accuracy ± 2.5 nm; precision 0.3 nm; noise ± 15 µAU @ 254 nm; drift 300 µAU/h; linearity > 2.0 AU @ 270 nm; max data rate 50 Hz (LAN); time constants 0.1-10 s; display matrix + keypad; flow cell not included.
  • Power Specifications: 100-240 V, 50-60 Hz, 75 W
  • Environmental Specifications: Temp 4-40 °C, humidity < 90 % non-condensing
  • Physical Specifications: 242 × 169 × 399 mm, 5.3 kg