Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: GD-MS
  • Manufacturer: Nu Instruments
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Available: Inactive
All specifications

About

Short description

Legacy high-resolution glow discharge mass spectrometer (GD-MS) optimized for ultra-trace impurity analysis in solid materials. Designed with cryogenic cooling, high-resolution adjustable slits, and robust ion optics for precise quantification across a wide dynamic range.

Specifications

Legacy high-resolution glow discharge mass spectrometer (GD-MS) optimized for ultra-trace impurity analysis in solid materials. Designed with cryogenic cooling, high-resolution adjustable slits, and robust ion optics for precise quantification across a wide dynamic range.

  • Status: Inactive
  • Country: United Kingdom of Great Britain and Northern Ireland
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Most elements between m/z 3 to 300, including trace and matrix-level components
  • Analyzer Type
    : Magnetic Sector MS
  • Ion Source Type: Glow discharge
  • Signal Stability / Relative Standard Deviation
    : <3% over 10 minutes; <5%over 30 minutes (for copper signals at RP>4000)
  • Inlet Types: Pin and flat sample holders for conductive and non-conductive materials, including powders and thin films
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : Adjustable resolution from 300 to over 10,000 (10% valley definition)
  • Sensitivity
    : 5×10−10 amps for copper signals at RP>4000
  • Dynamic Range
    : 12 orders of magnitude
  • Detection Limits
    : Sub-ppb for trace impurities in high-purity materials
  • Precision
    : Internal reproducibility: Major components <3%, Minor <5, Trace <10% RSD
  • Detector Type
    : Faraday detector with charge-coupled design and a secondary electron multiplier (SEM) for extended dynamic range
  • Discharge Parameters
    : Voltage: 1–1.2 kV; Current: 1–3.5 mA; Gas flow: 0.5–0.6 sccm
  • Geometry Type
    : Double-focusing with electrostatic and magnetic sectors
  • Focusing Type
    : Continuously adjustable zoom optics
  • Magnetic Sector Radius/ Angle
    : 25 cm radius, 90° deflection
  • Electric Sector Radius/ Angle
    : 32 cm radius/ 120 deg
  • Cooling System
    : Liquid nitrogen cryo-cooling and water chiller for turbomolecular pumps
  • Vacuum System
    : Three-zone vacuum system with turbo molecular and ion-getter pumps; operating pressure <2×10−7 mbar
  • Gas Supply: Argon (99.9995%) and compressed air
  • Power Specifications: 230 V, 50/60 Hz, 16 A; 1.5–2 kW consumption during operation
  • Environmental Specifications: Temperature: 20–24°C; Humidity: <70%
  • Physical Specifications: 1670 mm (W) × 1600 mm (D)

Other company product