Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Bruker
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Ion-trap LC/MS platform with design updates for proteomics and small-molecule analyses.

Specifications

Ion-trap LC/MS platform with design updates for proteomics and small-molecule analyses.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Ion Trap MS
  • High Resolution
    : No
  • Mass Resolution/ Mass Resolving Power
    : 0.3 u
  • Sensitivity
    : 250 fg reserpine: S/N > 500
  • Scan Modes
    : MS, MSⁿ
  • Ion Source Types [LC-MS]
    : ESI, APCI, CaptiveSpray, DIP
  • Scanning Speed
    : 52,000 u/sec
  • Polarity Switching
    : Zero Delay Alternating for polarity switching with 20 Hz MS acquisition speed
  • Fragmentation
    : CID, ETD, PTR
  • Vacuum System
    : air-cooled with turbopumps
  • Physical Specifications: 89.0 x 76.0 x 51.0 cm

Other company product