Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Triple quadrupole ICP-MS with MS/MS mode for effective interference removal.

Specifications

Triple quadrupole ICP-MS with MS/MS mode for effective interference removal.

  • Status: Active
  • Country: United States of America
  • Isotopic: Yes
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Analyzer Type
    : Tandem Quadrupole MS
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : <3% RSD over 20 minutes, <3% RSD over 2 hours (Factory Specifications)
  • High Resolution
    : Yes
  • Mass Resolution/ Mass Resolving Power
    : <0.3 amu (MS/MS mode)
  • Sensitivity
    : Varies by element, e.g., 150 Mcps/ppm for 7Li, 500 Mcps/ppm for 89Y
  • Dynamic Range
    : 10 orders for Standard; 11 orders for Advanced Applications
  • Detection Limits
    : <0.05 ppt for 115In; <0.02 ppt for 238U in no gas mode
  • Precision
    : <0.2% RSD for isotope ratio precision
  • Detector Type
    : Dual-mode discrete dynode electron multiplier
  • Oxides %
    : ≤1.5%
  • Reaction Cell System Details: ORS4 system with He, H2, NH3, O2, and other gases for advanced interference control
  • High Matrix Introduction System (HMI)
    : Yes
  • Spray Chamber
    : Scott-type double-pass, quartz, temperature range -5 °C to +20 °C
  • Nebulizer
    : MicroMist, borosilicate glass
  • Peristaltic Pump
    : 10-roller, 3 channels
  • Torch
    : One-piece quartz, 2.5 mm id injector with ShieldTorch system
  • Injector
    : 2.5 mm id
  • Interface
    : Sampling cone (Ni tip standard, optional Pt tip) and skimmer cone (Ni, optional Pt)
  • CRC Gas Flow
    : He, H2, NH3, O2, and other reactive/collision gases
  • Focusing Type
    : Off-axis Omega lens
  • Cooling System
    : Water chiller or water recirculator required
  • Vacuum System
    : Four-stage differential vacuum system with split-flow turbo molecular pumps
  • Gas Supply: Argon (99.99% purity), He (99.999%), NH3, O2 (various purities as required)
  • Reaction Cell System: Yes
  • Power Specifications: Single Phase, 200-240 V, 50/60 Hz, 30 A
  • Environmental Specifications: Temperature: 15–30 °C; Humidity: 20-80% non-condensing
  • Physical Specifications: Mainframe: 1060 mm width x 600 mm depth x 595 mm height; Weight: 139 kg