Specifications

  • Category: Inorganic and Isotopic Mass Spectrometry
  • Technique: ICP-MS
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

ICP-MS system for high-throughput analysis, handling up to 25% solids in samples.

Specifications

ICP-MS system for high-throughput analysis, handling up to 25% solids in samples.

  • Status: Active
  • Country: United States of America
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Analyzer Type
    : Quadrupole MS (single)
  • Ion Source Type: Inductively Coupled Plasma
  • Signal Stability / Relative Standard Deviation
    : <3% RSD over 20 minutes, <3% RSD over 2 hours (Factory Specifications)
  • Mass Resolution/ Mass Resolving Power
    : Variable from 0.3 u to 1.0 amu
  • Sensitivity
    : Varies by element, e.g., 110 Mcps/ppm for 7Li, 540 Mcps/ppm for 238U
  • Dynamic Range
    : 10 orders of magnitude (0.3 cps to 4 Gcps)
  • Detection Limits
    : <0.1 ppt for 115In, <0.04 ppt for 209Bi
  • Precision
    : <0.1% RSD for isotope ratio precision
  • Detector Type
    : Dual-mode discrete dynode electron multiplier
  • Oxides %
    : ≤1.5%
  • Reaction Cell System Details: Octopole Reaction System (ORS) with helium collision mode; optional H2 gas line
  • High Matrix Introduction System (HMI)
    : Yes
  • Spray Chamber
    : Scott-type double-pass, quartz, temperature range -5 °C to +20 °C
  • Nebulizer
    : MicroMist, borosilicate glass
  • Peristaltic Pump
    : 10-roller, 3 channels
  • Torch
    : One-piece quartz, 2.5 mm id injector with ShieldTorch system
  • Injector
    : 2.5 mm id
  • Interface
    : Sampling cone (Ni-tipped standard, optional Pt-tipped) and skimmer cone (Ni, optional Pt-tipped)
  • CRC Gas Flow
    : Helium and optional hydrogen
  • Focusing Type
    : Off-axis Omega lens
  • Cooling System
    : Water chiller or water recirculator required
  • Vacuum System
    : Three-stage differential vacuum system with turbo molecular and rotary pumps
  • Gas Supply: Argon (99.99% purity), Helium (99.999% purity), H2 (99.999% purity)
  • Reaction Cell System: Yes
  • Power Specifications: Single Phase, 200-240 V, 50/60 Hz, 30 A
  • Environmental Specifications: Temperature: 15–30 °C; Humidity: 20-80% non-condensing
  • Physical Specifications: Mainframe: 730 mm width x 600 mm depth x 595 mm height; Weight: 100 kg

Other company product