Specifications

  • Category: Chromatography
  • Technique: GC
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

GC for routine analysis. Supports EPC and EPR for precise control.

Specifications

GC for routine analysis. Supports EPC and EPR for precise control.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Column Type [GC]
    : Capillary, Packed
  • Available Detectors [GC]
    : Flame Ionization (FID), Thermal Conductivity (TCD), Electron Capture (ECD), Nitrogen-Phosphorus (NPD), Flame Photometric (FPD)
  • FID Details
    : Max Temp: 425°C, MDL: <3 pg C/s, Linear range: 10^7
  • TCD Details
    : Max Temp: 400°C, MDL: <800 pg/mL
  • ECD Details
    : Max Temp: 400°C, MDL: <0.02 pg/mL lindane
  • NPD Details
    : Max Temp: 400°C, MDL: <0.4 pg N/s, <0.2 pg P/s
  • FPD Details
    : Max Temp: 250°C, MDL: <200 fg P/s, <6 pg S/s, Selectivity: 106 g S/g C, 106 g P/g C
  • Inlet Types [GC]
    : Split/Splitless , Purged Packed, PCOC
  • Injector General Parameters
    : Max Temp: 400°C, Pressure: 0-100 psi
  • Flow Control
    : Electronic (EPC and EPR)
  • Autosampler Compatibility: Yes (7693A autoinjector)
  • Data Acquisition And Control: LAN, Analog, RS-232
  • Column Oven Specifications
    : 8°C above ambient to 425°C, 5 ramps
  • Safety Features
    : Over-temperature protection
  • Power Specifications: 100-240 V, 1500-2250 W
  • Environmental Specifications: Ambient: 15-30°C, Humidity: 30-70%
  • Physical Specifications: H: 49 cm, W: 56 cm, D: 51 cm, 50 kg