Instruments
Agilent 7800
Main
Inst. Database
Inorganic and Isotopic Mass Spectrometry
Inductively Coupled Plasma Mass Spectrometry
Single Quad ICP-MS
Agilent 7800
Specifications
Category:
Inorganic and Isotopic Mass Spectrometry
Technique:
ICP-MS
Manufacturer:
Agilent
Country:
United States of America
Available:
Inactive
All specifications
Download "Conclusions.PDF"
Download "Conclusions.PDF"
About
Specifications
About
Short description
Retired quadrupole ICP-MS.
All about
Specifications
Retired quadrupole ICP-MS.
Status
:
Inactive
Country
:
United States of America
Elemental
:
Yes
Analyte Type / Target Of Analysis
:
Metals and non-metals
Analyzer Type
:
Quadrupole MS (single)
Ion Source Type
:
Inductively Coupled Plasma
Signal Stability / Relative Standard Deviation
:
<3% RSD over 20 minutes, <3% RSD over 2 hours (Factory Specifications)
Mass Resolution/ Mass Resolving Power
:
Variable from 0.3 u to 1.0 amu
Sensitivity
:
Varies by element, e.g., 110 Mcps/ppm for 7Li, 540 Mcps/ppm for 238U
Dynamic Range
:
10 orders of magnitude (0.1 cps to 4 Gcps)
Detection Limits
:
<0.1 ppt for 115In, <0.04 ppt for 209Bi
Precision
:
<0.1% RSD for isotope ratio precision
Detector Type
:
Dual-mode discrete dynode electron multiplier
Oxides %
:
CeO/Ce <1.5%
Reaction Cell System Details
:
Octopole Reaction System (ORS) with helium collision mode; optional H2 gas line
High Matrix Introduction System (HMI)
:
Yes
Spray Chamber
:
Scott-type double-pass, quartz, temperature range -5 °C to +20 °C
Nebulizer
:
MicroMist, borosilicate glass
Peristaltic Pump
:
10-roller, 3 channels
Torch
:
One-piece quartz, 2.5 mm id injector with ShieldTorch system
Injector
:
2.5 mm id
Interface
:
Sampling cone (Ni-tipped standard, optional Pt-tipped) and skimmer cone (Ni, optional Pt-tipped)
CRC Gas Flow
:
Helium and optional hydrogen
Focusing Type
:
Off-axis Omega lens
Cooling System
:
Water chiller or water recirculator required
Vacuum System
:
Three-stage differential vacuum system with turbo molecular and rotary pumps
Gas Supply
:
Argon (99.99% purity), Helium (99.999% purity), H2 (99.999% purity)
Reaction Cell System
:
Yes
Power Specifications
:
Single Phase, 200-240 V, 50/60 Hz, 30 A
Environmental Specifications
:
Temperature: 15–30 °C; Humidity: 20-80% non-condensing
Physical Specifications
:
Mainframe: 730 mm width x 600 mm depth x 595 mm height; Weight: 100 kg