Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Quadrupole-TOF GC-MS/MS suitable for complex matrix analysis, enhanced molecular ion info and suitable for structural elucidation.

Specifications

Quadrupole-TOF GC-MS/MS suitable for complex matrix analysis, enhanced molecular ion info and suitable for structural elucidation.

  • Status: Active
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Quadrupole MS (single), Time-of-Flight MS
  • Mass Resolution/ Mass Resolving Power
    : Selectable, 0.7 to 3.0 Daltons amusing default tune Settable, 0.4 to 4.0 Daltons amusing custom tune, TOF >25,000 at m/z 271.9867
  • Sensitivity
    : < 60 fg OFN (EI mode)
  • Dynamic Range
    : > 10⁵
  • Available Configurations: EI
  • Scan Modes
    : Full scan, SIM,MS/MS
  • Ionization Modes [GC-MS]
    : Electron (EI)
  • Mass Accuracy/Stability
    : < 2 ppm RMS
  • Detector Type [GC-MS]
    : Microchannel Plate / PMT
  • Scanning Speed
    : 1–50 Hz
  • Ion Source Temperature
    : 100 to 350°C
  • Collision Cell
    : Linear hexapole (4:1 He:N2)
  • Installation Specifications
    : EI IDL <60 fg (QTOF); TOF Accuracy <5 ppm
  • Vacuum System
    : Four-stage turbo pump (200/200 L/sec N2 and two 300 L/sec N2)