Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: GC-MS
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired Quadrupole-TOF GC-MS/MS suitable for complex matrix analysis.

Specifications

Retired Quadrupole-TOF GC-MS/MS suitable for complex matrix analysis.

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (volatile and semi-volatile)
  • Analyzer Type
    : Quadrupole MS (single), Time-of-Flight MS
  • Mass Resolution/ Mass Resolving Power
    : Selectable, 0.7 to 3.0 Daltons amusing default tune Settable, 0.4 to 4.0 Daltons amusing custom tune
  • Sensitivity
    : < 250 fg OFN (EI mode)
  • Dynamic Range
    : > 10⁵
  • Available Configurations: EI, EI/CI
  • Scan Modes
    : Full scan, SIM,MS/MS
  • Ionization Modes [GC-MS]
    : Electron (EI), Chemical Ionization (CI), Positive/Negative Chemical Ionization (PCI/NCI)
  • Mass Accuracy/Stability
    : < 5 ppm (typically < 2 ppm)
  • Detector Type [GC-MS]
    : Microchannel Plate / PMT
  • Scanning Speed
    : 1–50 spectra/sec
  • Ion Source Temperature
    : 106 to 350°C
  • Maximum Column Flow Rate
    : Up to 8 mL/min
  • Collision Cell
    : Linear hexapole
  • Installation Specifications
    : EI IDL: <250 fg (99% confidence, 8 splitless injections of 1 pg OFN, m/z 271.9867, <8% RSD) EI SNR: >2000:1 (RMS noise, full spectrum, best of 8 injections, 1 pg OFN, m/z 271.9867) EI TOF Resolution: >12,500 (FWHM, 1 pg OFN, m/z 271.9867; typically >13,500) EI TOF Accuracy: <5 ppm (1 pg OFN, m/z 271.9867 ± 0.0014; typically <2 ppm) PCI SNR: >1500:1 (RMS noise, 100 pg BZP, methane reagent gas, m/z 183.0804)
  • Vacuum System
    : Four-stage turbo pump (200/200 L/sec N2 and two 300 L/sec N2)
  • Physical Specifications: MS only: 63.5×89×47 cm, 148 kg