Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired ion trap LC/MS system with data-dependent MSⁿ and support for electron transfer dissociation (ETD).

Specifications

Retired ion trap LC/MS system with data-dependent MSⁿ and support for electron transfer dissociation (ETD).

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Ion Trap MS
  • High Resolution
    : No
  • Mass Resolution/ Mass Resolving Power
    : ≤ 0.6 u to ≤ 0.25 u depending on scan speed
  • Sensitivity
    : Reserpine on-column: 5 pg (≥25:1), 1 pg (≥50:1), 250 fg (≥50:1); 6340: TBD
  • Scan Modes
    : Full scan MS/MS, data-dependent MSⁿ
  • Ion Source Types [LC-MS]
    : ESI, APCI, APPI, Multimode (ESI+APCI), HPLC-Chip, AP/MALDI PDF
  • Mass Accuracy/Stability
    : ±0.2 u; Stability: ±0.2 u over 8 hours
  • Scanning Speed
    : 27,000 u/s (maximum, depending on resolution setting)
  • Polarity Switching
    : ≈ 0.5 s
  • Fragmentation
    : CID, ETD, ETD+CID

Other company product