Specifications

  • Category: Spectroscopy
  • Technique: ICP-OES
  • Manufacturer: Agilent
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Retired dual-view ICP-OES analyzer for simultaneous axial and radial measurements. Incorporates Dichroic Spectral Combiner (DSC) for the fastest analysis with low gas usage.

Specifications

Retired dual-view ICP-OES analyzer for simultaneous axial and radial measurements. Incorporates Dichroic Spectral Combiner (DSC) for the fastest analysis with low gas usage.

  • Status: Active
  • Country: United States of America
  • Elemental: Yes
  • Analyte Type / Target Of Analysis: Metals and non-metals
  • Measurable Isotopes/Elements: Elemental coverage including trace and major components
  • Signal Stability / Relative Standard Deviation
    : Long-term stability with <1% RSD over 8 hours without internal standardization
  • Sensitivity
    : Sub-ppb detection limits for trace elements
  • Dynamic Range
    : Linear dynamic range of 6–9 orders of magnitude, depending on configuration
  • Spray Chamber
    : Glass cyclonic double-pass spray chamber; optional configurations for HF-resistant and high-salt applications
  • Nebulizer
    : Glass concentric nebulizer; optional high-sensitivity and HF-resistant nebulizers available
  • Peristaltic Pump
    : Fully PC-controlled peristaltic pump with variable speeds (0–80 rpm), up to five channels
  • View Orientation/Observation/Optical System
    : Vertical torch with SVDV, radial, and axial configurations; Dichroic Spectral Combiner (DSC) technology for simultaneous radial and axial viewing
  • Detection System/Detector
    : VistaChip II CCD detector
  • Grating /Wavelength Range
    : 167–785 nm with echelle grating; single slit design for high resolution
  • Polychromator
    : Thermostatted polychromator (35°C) with high optical stability; CaF₂ prism cross-disperser
  • Typical Resolution
    : <7 pm (UV region); optimized for both trace and major element measurements
  • Gas Supply: Argon ≥99.996% purity; mass flow-controlled plasma, auxiliary, nebulizer, and makeup gas flows
  • Power Specifications: 200–240 VAC, 50/60 Hz; ~2.9 kVA
  • Environmental Specifications: Operating temperature: 15–35°C; Humidity: 20–80% (non-condensing); Exhaust flow: 2.5 m³/min
  • Physical Specifications: 800 mm W × 740 mm D × 940 mm H; Weight: 106 kg

Other company product