Specifications

  • Category: Chromatography
  • Technique: Unknown
  • Manufacturer: OI Analytical
  • Country: United States of America
  • Available: Active
All specifications

About

Short description

Pulsed Flame Photometric Detector for sulfur and phosphorus detection in gas chromatography. Provides up to tenfold higher sensitivity than conventional FPDs with self-cleaning flame design. Offers modular electronics and flow modules for reliable, low-maintenance operation.

Specifications

Pulsed Flame Photometric Detector for sulfur and phosphorus detection in gas chromatography. Provides up to tenfold higher sensitivity than conventional FPDs with self-cleaning flame design. Offers modular electronics and flow modules for reliable, low-maintenance operation.

  • Status: Active
  • Country: United States of America
  • Device Class (Function)
    : Ionization / Flame Detector
  • Technique Compatibility
    : GC
  • Brand Independence
    : Yes
  • Performance Metrics
    : Detectivity: S <1 pg S/s; P <100 fg P/s. Sensitivity: S/N >300 (10 pg S/s). Drift: <10× noise in 20 min. Selectivity: S >10⁶ S/C; P >10⁵ P/C (adjustable). Linearity: S quadratic, 3 decades dynamic range; P linear, 5 decades. Equimolar ±8%. Min temp 180 °C, max 420 °C. Gases: He/H₂ (40 psig), air (40 psig), H₂ (40 psig). Gas use: H₂ 10-15 mL/min, air 20-30 mL/min.
  • Power Specifications: 115/230 VAC
  • Environmental Specifications: 5-80 % RH, altitude ≤2000 m
  • Physical Specifications: Controller: 17.5 × 6 × 25 cm; Pneumatic module: 17.5 × 6 × 27.5 cm