Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Varian
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired triple quadrupole LC/MS; vESI ion source; Gold Guard ion optics

Specifications

Retired triple quadrupole LC/MS; vESI ion source; Gold Guard ion optics

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Tandem Quadrupole MS
  • High Resolution
    : No
  • Mass Resolution/ Mass Resolving Power
    : Unit mass resolution (0.7 amu, FWHM)
  • Sensitivity
    : vESI: 100 fg Reserpine S/N ≥800:1; APCI: 2 pg Reserpine S/N ≥150:1
  • Dynamic Range
    : 6 orders of magnitude
  • Scan Modes
    : SIM (Q1/Q3), Full scan (Q1/Q3), Precursor ion scan, Product ion scan, Neutral loss scan, SRM, MRM
  • Ion Source Types [LC-MS]
    : vESI (standard), APCI (optional)
  • Mass Accuracy/Stability
    : ±0.1 amu over 24 h
  • Scanning Speed
    : Up to 6000 amu/sec
  • Fragmentation
    : CID via 180° lensless square-geometry collision cell
  • Vacuum System
    : Two-stage turbomolecular pump (310/400 L/sec) + 2 MS40+ foreline pumps
  • Power Specifications: MS: 100-240 Vac, 50/60 Hz ±3 Hz, 1.0 kW; MS40+ foreline: 200-240 Vac, 50/60 Hz, 1.2 kW each
  • Environmental Specifications: Humidity: 20-80% RH (non-condensing); Temp: 16-30 °C
  • Physical Specifications: 50.0 × 38.0 × 71.0 cm; 108.0 kg