Specifications

  • Category: Organic Mass-Spectrometry
  • Technique: LC-MS
  • Manufacturer: Varian
  • Country: United States of America
  • Available: Inactive
All specifications

About

Short description

Retired LC/MS with upgradeable single to triple quadrupole; Flexible API/GC interface; Broad scan and SRM modes

Specifications

Retired LC/MS with upgradeable single to triple quadrupole; Flexible API/GC interface; Broad scan and SRM modes

  • Status: Inactive
  • Country: United States of America
  • Analyte Type / Target Of Analysis: Organic molecules (non-volatile, polar)
  • Analyzer Type
    : Quadrupole MS (single), Tandem Quadrupole MS
  • High Resolution
    : No
  • Mass Resolution/ Mass Resolving Power
    : Unit mass resolution
  • Sensitivity
    : +ESI SIM: 500 fg Reserpine ≥10:1; +ESI MS/MS: 500 fg ≥20:1
  • Dynamic Range
    : 5 orders of magnitude
  • Scan Modes
    : Full scan, SIM, Precursor scan, Product scan, Neutral loss scan, SRM
  • Ion Source Types [LC-MS]
    : ESI (standard), APCI (optional), GC interface (optional)
  • Mass Accuracy/Stability
    : ±0.1 Da over 24 h
  • Scanning Speed
    : Up to 6000 Da/sec
  • Polarity Switching
    : 170 ms
  • Fragmentation
    : CID via 180° curved collision cell
  • Vacuum System
    : Dual-stage turbomolecular pump (310/400 L/sec) + 2 foreline pumps
  • Power Specifications: MS: 100-240 Vac, 50/60 Hz ±3 Hz, 1.0 kW; Dual foreline pumps: 100-240 Vac, 50/60 Hz, 5.5 kW
  • Environmental Specifications: Humidity: 20-80% RH (non-condensing); Temp: 16-30 °C
  • Physical Specifications: 50.0 × 38.0 × 71.0 cm; 108.0 kg

Other company product